Wide and narrow disposable blades can be used, convenient and simple rotary slicer can directly use disposable blades, wide and narrow disposable blades can be used, convenient and simple. The streamlined cover is neat and tidy. The top of the cover can be placed with blades and wax blocks. The imported roller guides and high-precision roller screw are used. The design is reasonable, the structure is tight, the precision is high, and the stability is reliable. Ideal product.
1. Installation of rotary slicer
Place the instrument where the following conditions are met:
The test bench is stable and has no vibration.
The ground should be free of vibration.
The instrument operation should have a spacious and comfortable operating space.
The room temperature is maintained between +5 ° C and +40 ° C.
WARNING: Do not hold the handwheel, tool holder, wax block pliers or slice thickness adjustment rotation when handling the instrument!
2, clip specimen
Shake the movable block to the highest position and lock the handwheel to rotate the flat handle on the specimen holder to adjust the size of the specimen holder.
Place the specimen in the desired position.
Rotate the clamp handle on the specimen holder to clamp the specimen.
Turn the wax block retainer lock wrench clockwise to adjust the surface on which the specimen has been clamped.
Turn the wax block retainer locking wrench counterclockwise to lock the specimen holder.
Warning: The specimen should be clamped before the knife is installed! Always lock the handwheel each time you operate the knife and specimen or at work.
3, with a knife
First install the left and right guard covers and turn the left and right slicing knife fixing knobs counterclockwise to make them exit.
Loosen the tilt angle adjustment wrench of the slicing knife.
Hold the knife back of the knife with the blade facing up and carefully insert the knife holder from the side.
Turn the blade fixing knob clockwise to evenly resist the cutting blade.
Move the blade tilt angle adjustment wrench to adjust the rear corner of the cutting blade to the desired position, and then tighten the cutter blade tilt angle to adjust the wrench bolt.
Turn the blade fixing knob clockwise to evenly clamp the cutting blade.
Warning: Be very careful when using the slicing knife. The cutting edge is very sharp and mishandling will cause serious injury!
4, rough cut repair block
Pull the handwheel to stop rotating the handle and release the specimen holder. Turn the handwheel to shake the specimen to the opposite position of the knife.
Release the clutch, move the knife holder to bring the blade close to the specimen, and then lock the clutch.
Turn the slice thickness thin and coarse adjuster to adjust the specimen to just meet the slicing knife. At the same time, turn the handwheel to move the specimen up and down, and observe the distance between the specimen and the blade, and flatten the specimen wax block.
Turn the slice thickness spinner to the thickness (eg 20um) and turn the handwheel to test the specimen until the desired specimen surface.
5, slice
Adjust the slice thickness to the desired value by turning the slice thickness spinner.
Turn the handwheel evenly to slice.
Industrial electric furnace http://
Emulsified asphalt equipment http://
Http://news.chinawj.com.cn Editor: (Hardware Business Network Information Center) http://news.chinawj.com.cn
1. Installation of rotary slicer
Place the instrument where the following conditions are met:
The test bench is stable and has no vibration.
The ground should be free of vibration.
The instrument operation should have a spacious and comfortable operating space.
The room temperature is maintained between +5 ° C and +40 ° C.
WARNING: Do not hold the handwheel, tool holder, wax block pliers or slice thickness adjustment rotation when handling the instrument!
2, clip specimen
Shake the movable block to the highest position and lock the handwheel to rotate the flat handle on the specimen holder to adjust the size of the specimen holder.
Place the specimen in the desired position.
Rotate the clamp handle on the specimen holder to clamp the specimen.
Turn the wax block retainer lock wrench clockwise to adjust the surface on which the specimen has been clamped.
Turn the wax block retainer locking wrench counterclockwise to lock the specimen holder.
Warning: The specimen should be clamped before the knife is installed! Always lock the handwheel each time you operate the knife and specimen or at work.
3, with a knife
First install the left and right guard covers and turn the left and right slicing knife fixing knobs counterclockwise to make them exit.
Loosen the tilt angle adjustment wrench of the slicing knife.
Hold the knife back of the knife with the blade facing up and carefully insert the knife holder from the side.
Turn the blade fixing knob clockwise to evenly resist the cutting blade.
Move the blade tilt angle adjustment wrench to adjust the rear corner of the cutting blade to the desired position, and then tighten the cutter blade tilt angle to adjust the wrench bolt.
Turn the blade fixing knob clockwise to evenly clamp the cutting blade.
Warning: Be very careful when using the slicing knife. The cutting edge is very sharp and mishandling will cause serious injury!
4, rough cut repair block
Pull the handwheel to stop rotating the handle and release the specimen holder. Turn the handwheel to shake the specimen to the opposite position of the knife.
Release the clutch, move the knife holder to bring the blade close to the specimen, and then lock the clutch.
Turn the slice thickness thin and coarse adjuster to adjust the specimen to just meet the slicing knife. At the same time, turn the handwheel to move the specimen up and down, and observe the distance between the specimen and the blade, and flatten the specimen wax block.
Turn the slice thickness spinner to the thickness (eg 20um) and turn the handwheel to test the specimen until the desired specimen surface.
5, slice
Adjust the slice thickness to the desired value by turning the slice thickness spinner.
Turn the handwheel evenly to slice.
Industrial electric furnace http://
Emulsified asphalt equipment http://
Http://news.chinawj.com.cn Editor: (Hardware Business Network Information Center) http://news.chinawj.com.cn
Applicable to medical equipment and beds
Mute effect: Bearing Steel mute, high strength, low noise
Humanized design, stable braking performance, Easier to use.
Medical Caster Wheels,Medical Casters,hospital castor
Yangjiang Xingyang Industry & Trade Co.,Ltd. , https://www.xycaster.com